GENERAL SPECIFICATIONS @ 68 °F
| .Output |
.4 to 20 mA loop powered,
Max. 30 ma |
| .Loop Supply |
.10 to 30 VDC reverse
connection protected |
| .Loop Resistance |
.700 ohms @ 24 V |
| .Loop Volts
Sensitivity |
.0.05%/V |
| .Temperature
Stability |
.ZERO drift = 0.01%/°F |
| .SPAN = 0.002%/°F |
| .Ambient
Temperature |
.32 to 150 °F at 24 V
supply 250 ohm load |
| .Ambient
RH |
.0 to 95% RH non condensing |
| .EMC
Compatibility |
.Emissions = BS.EN 50081 |
| .Susceptibility = BS.EN
50082 |
INPUTS
| .SEM104P
PLATINUM RESISTANCE SENSOR |
| .Sensor |
.100 ohm - 3 wire - alpha =
0.00385 ohms |
| .Accuracy |
.±0.4 °F plus +0.2% of
reading |
| .Offset adjustment |
.range dependant |
| .Span adjustment |
.76 °F to 932 °F |
| .SEM104K,
T or J THERMOCOUPLE SENSOR (ISOLATED JUNCTION) |
| .Accuracy |
.±0.1% FS plus cold
junction errors non-linearized |
| .Offset adjustment |
.±200 °F, solder link
& potentiometer |
| .Span adjustment |
.212 to 1832 °F |
| .Impedance |
.>1M ohm |
| .Burn Out |
.Upscale standard,
Downscale by link |
| .Cold Junction |
.Automatic 32 to 158 °F |
| |
.Accuracy 0.4 °F @ 68 °F,
tracking +0.02 °F/°F |
| .Thermocouple Types |
.J = Iron
/ Constantan |
| |
.T = Chromel
/ Constantan |
| |
.K =
Chromel / Alumel |
STANDARD RANGES °F
| Range °F |
SEM104P |
|
SEM104K |
SEM104T |
SEM104J |
| -40 to +140 |
|
|
* |
* |
* |
| -100 to +100 |
* |
|
|
* |
|
| 0 to 200 |
* |
|
* |
* |
* |
| 0 to 300 |
* |
|
* |
* |
* |
| 0 to 500 |
* |
|
* |
* |
* |
| 0 to 750 |
* |
|
* |
* |
* |
| 0 to 1000 |
* |
|
* |
|
* |
| 0 to 1500 |
* |
|
|
|
* |
STANDARD RANGES °C
| Range °C |
SEM104P |
|
SEM104K |
SEM104T |
SEM104J |
| -30 to +35 |
* |
|
|
|
|
| -25 to +75 |
* |
|
|
|
|
| 0 to 50 |
* |
|
|
|
|
| 0 to 100 |
* |
|
* |
* |
* |
| 0 to 150 |
* |
|
* |
* |
* |
| 0 to 200 |
* |
|
* |
* |
* |
| 0 to 250 |
* |
|
* |
* |
* |
| 0 to 400 |
* |
|
* |
* |
* |
| 0 to 500 |
* |
|
* |
|
* |
| 0 to 600 |
|
|
* |
|
|
| 0 to 800 |
|
|
* |
|
|
| 0 to 1000 |
|
|
* |
|
|
Other ranges can be supplied factory set or they can be configured by the use
of solder links on the SEM pcb. Full details are supplied with each
transmitter.
DOWNLOADS
SEM104P Instruction Manual
SEM104TC Instruction Manual
SEM104 Literature
ORDER CODE
SEM104 / SENSOR /
Lo RANGE / Hi RANGE / °F or °C
Example:
SEM104/K/0/200/F is the code for an SEM104
with type K thermocouple input with 4 to 20 ma output over the range of 0 to 200
°F